User Contributed MET/CAL PROCEDURE ============================================================================== INSTRUMENT: Triplett 310 DATE: 15-Nov-88 AUTHOR: User Contributed REVISION: V 4.0 ADJUSTMENT THRESHOLD: 100% NUMBER OF TESTS: 20 NUMBER OF LINES: 79 CONFIGURATION: Fluke 5100B CONFIGURATION: Fluke 5205A ============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP Zero the Meter Movement using the Center Screw. 1.001 DISP (The Triplett should be laying on its back) 1.002 ASK- F 1.003 DISP Connect the meter probes into the 'Binding Post 1.003 DISP to Double Banana Plug Adapter' Red to Red and Black 1.003 DISP to Black. Insert the leads from the right when the 1.003 DISP adapter Red terminal is to the top. 1.004 DISP Insert the Black lead in the meter COM input and the 1.004 DISP Red lead in the VOM input. 1.005 HEAD {Volts DC and Linearity Tests} 1.006 5100 3 1V 3/ 2W 2.001 5100 3 2V 3/ 2W 3.001 5100 3 3V 3/ 2W 4.001 5100 12 12V 3/ 2W 5.001 5100 60 60V 3/ 2W 6.001 5100 300 300V 3/ 2W 7.001 DISP Remove the Red wire from the Test Instrument VOM input 7.001 DISP and connect to the 1200 VDC input. 7.002 ASK- R 7.003 5205 1200 1000V 3/ 2W 8.001 HEAD {Volts AC Tests} 8.002 DISP Remove the Red wire from the Test Instrument 1200 VDC 8.002 DISP input and connect to the 1200 VAC input. 8.003 DISP Set the meter to the 300 Volts AC Range. 8.004 5205 1200 1000V 4/ 400H 2W 9.001 DISP Remove the Red wire from the Test Instrument 1200 VAC 9.001 DISP input and connect to the VOM input. 9.002 5100 300 300V 4/ 400H 2W 10.001 ASK+ R 10.002 5100 60 60V 4/ 400H 2W 11.001 5100 12 12V 4/ 400H 2W 12.001 HEAD {Current DC Tests} 12.002 5100 0.6 0.6mA 3/ 2W 13.001 5100 6 6mA 3/ 2W 14.001 5100 60 60mA 3/ 2W 15.001 5100 600 600mA 3/ 2W 16.001 HEAD {Resistance Tests} 16.002 ASK- R N 16.003 MESS Adjust the System for 16.003 MESS 16.003 MESS 150 indicated on 0 to 300-volt dc scale. 16.004 DISP Set Range to {X1} OHMS. 16.005 DISP Short the wires from the Test Instrument together at 16.005 DISP the Test Interface Panel. 16.005 DISP Adjust the ohms control for 0 reading on the meter. 16.005 DISP Remove the short. 16.006 RESF 100Z 12.5% -11% 2W 17.001 DISP Set Range to {X10} OHMS. 17.002 DISP Short the wires from the Test Instrument together at 17.002 DISP the Test Interface Panel. 17.002 DISP Adjust the ohms control for 0 reading on the meter. 17.002 DISP Remove the short. 17.003 RESF 1000Z +12.5% -11% 2W 18.001 DISP Set Range to {X100} OHMS. 18.002 DISP Short the wires from the Test Instrument together at 18.002 DISP the Test Interface Panel. 18.002 DISP Adjust the ohms control for 0 reading on the meter. 18.002 DISP Remove the short. 18.003 RESF 10000Z +12.5% -11% 2W 19.001 DISP Set Range to {X1K} OHMS. 19.002 DISP Short the wires from the Test Instrument together at 19.002 DISP the Test Interface Panel. 19.002 DISP Adjust the ohms control for 0 reading on the meter. 19.002 DISP Remove the short. 19.003 RESF 100KZ +12.5% -11% 2W 20.001 END